Masimo Signal Extraction Technology (SET) pulse oximetry begins with conventional red and infrared photoplethysmographic signals, and then employs a constellation of advanced techniques including radiofrequency and light-shielded optical sensors, digital signal processing, and adaptive filtration, to measure SpO2 accurately during challenging clinical conditions. In contrast to CPO which calculates O2 saturation from the ratio of transmitted pulsatile red and infrared light, Masimo SET pulse oximetry uses a new conceptual model of light absorption for pulse oximetry and employs the discrete saturation transform (DST) to isolate individual "saturation components" in the optical pathway. Typically, when the tissue under analysis is stationary, only the single saturation component produced by pulsatile arterial blood is present. In contrast, during patient motion, movement of non-arterial components (for example, venous blood) can be identified as additional saturation components (with a lower O2 saturation). When conditions of the Masimo model are met, the saturation component corresponding to the highest O2 saturation is reported by the instrument as SpO2.